Multi 350i
Operation
29
4.4.4
Settings for pH and ORP measurements
Overview
The following settings are possible for pH and ORP measurements:
!
Resolution
!
Calibration interval
!
Buffers for calibration
!
Unit for slope
!
Calibration record
(display)
Settings
The settings are made in the measuring menu of the pH/ORP measure-
ment. To open the settings, activate the relevant measuring window in
the measured value display and press the
<MENU/ENTER>
key short-
ly. After completing the settings, switch to the measured value display
with
<M>
.
Menu item
Possible
setting
Description
Calibration
/
Calibration inter-
val
1 ... 999 d
Calibration interval
for the pH elec-
trode (in days).
The measuring instrument reminds
you to calibrate regularly by the
flashing sensor symbol in the mea-
suring window.
Calibration
/
Calibration type
TEC
NIST/DIN
ConCal
Buffer sets to be used for pH cali-
bration.
For details, see section 4.4.5.
Calibration
/
Unit
for slope
mV/pH
%
Unit of the slope.
The % display refers to the Nernst
slope of
-59.16 mV/pH (100 x determined
slope/Nernst slope).
Calibration
/
Calibration
record
-
Displays the calibration record of
the last calibration.
Man. tempera-
ture
-20 ...
+130 °C
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Reset
-
Resets all sensor settings to the
delivery condition (see
section 4.10.1).
High resolution
On
Off
Resolution of the pH display:
On
= 0.001
Off
= 0.01