Multi 3430
Operation
67
ba75865e02
12/2009
4.8.4
Settings for IDS conductivity sensors
Overview
The following settings are possible for IDS conductivity sensors:
z
Calibration record
(display)
z
Calibration interval
z
Reference temperature
z
Temperature compensation
z
TDS factor
z
Unit of the temperature
z
Automatic
Stability control
Settings
The settings are made in the
Measurement
menu for the measured
parameter, conductivity. To open the settings, display the required
measured parameter in the measured value display and press the
<MENU/ENTER>
key. After completing the settings, switch to the mea-
sured value display with
<M>
.
The possible settings are individually displayed for each sensor. Below
the setting menu is displayed for two IDS sensors (TetraCon 925,
LR325/01).
Default settings are printed in
bold
.
Setting menu of
TetraCon 925
Menu item
Possible
setting
Description
Calibration
/
Cali-
bration record
-
Displays the calibration record of
the last calibration.
Calibration
/
Cali-
bration interval
1 ...
150
...
999 d
Calibration interval
for the IDS con-
ductivity sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring window.
Measurement
/
Measuring cell
/
Type
Measuring cell
used
cal
Measuring cells whose cell con-
stant is determined by calibration
in the KCL control standard solu-
tion.
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant is
displayed in the status line.
man
Freely adjustable cell constant in
the range 0.450 ... 0.500 cm
-1
.
Содержание Multi 3430
Страница 6: ...Contents Multi 3430 6 ba75865e01 12 2009...
Страница 14: ...Overview Multi 3430 14 ba75865e02 12 2009...
Страница 18: ...Safety Multi 3430 18 ba75865e02 12 2009...
Страница 22: ...Commissioning Multi 3430 22 ba75865e02 12 2009...
Страница 86: ...Operation Multi 3430 86 ba75865e02 12 2009...
Страница 100: ...Appendix Firmware update Multi 3430 100 ba75865e02 12 2009...
Страница 101: ......