4.4.3 Extended access control with instrument block
If the extended access control with instrument block is enabled, only
the IQ S
ENSOR
N
ET
logo is displayed on the terminal where the instru
-
ment block was activated.
As soon as a password was entered, a symbol for the currently enabled
access authorization is displayed.
Crown: Administrator logged in, all functions allowed
Tool: User with authorization for maintenance activities logged in (e.g.
calibration)
Eye: Observer without authorization
The relevant symbol is displayed depending on the currently active
user authorization. Functions that are not allowed are not displayed in
the menu.
Access as observer, user
or administrator
User authorization is required for all jobs, even for viewing the mea
-
sured values. You have to log in with a password. This can be done as
follows:
Press the any button.
The password prompt appears.
Enter the password with the arrow keys and confirm with OK.
The relevant user authorization is enabled.
Connect the electronic key to the USB interface of the MIQ/
TC 2020 XT terminal
(if the administrator has enabled this function)
The relevant user authorization is enabled.
Logging off
Protecting the system and
terminal
There are the following ways to protect the system and to lock the ter
-
minal:
Select the
Display/Options
/
Lock Device
menu.
The system is protected. The instrument is blocked.
Disconnect the electronic key from the USB interface of the MIQ/
TC 2020 XT terminal.
After 10 minutes without keystroke, the user or administrator is auto
-
matically logged off from the system and the system safety is auto
-
matically activated.
Содержание IQ SENSOR NET System 2020 XT
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 8: ...Contents System 2020 XT CONNECT 0 6 ba75970e01 10 2012...
Страница 24: ...Overview System 2020 XT CONNECT 1 16 ba75970e01 10 2012...
Страница 28: ...Safety System 2020 XT CONNECT 2 4 ba75970e01 10 2012...
Страница 123: ...System 2020 XT CONNECT Settings setup 5 11 ba75970e01 10 2012 Fig 5 10 Edit list of sensors...
Страница 138: ...Settings setup System 2020 XT CONNECT 5 26 ba75970e01 10 2012...
Страница 152: ...Network connection System 2020 XT CONNECT 6 14 ba75970e01 10 2012...
Страница 178: ...PROFIBUS connection option System 2020 XT CONNECT 8 12 ba75970e01 10 2012...
Страница 180: ...Maintenance and cleaning System 2020 XT CONNECT 9 2 ba75970e01 10 2012...
Страница 194: ...What to do if System 2020 XT CONNECT 10 14 ba75970e01 10 2012...
Страница 204: ...Accessories and options System 2020 XT CONNECT 12 2 ba75970e01 10 2012...
Страница 211: ...Lists IQ SENSOR NET System 2020 XT CONNECT 13 7 ba75970e01 10 2012 W Weekly load diagram of selected sensor 4 19...
Страница 212: ...Lists IQ SENSOR NET System 2020 XT CONNECT 13 8 ba75970e01 10 2012...
Страница 215: ......