System 2020 XT C
ONNECT
Operation
4 - 27
ba75970e01
10/2012
4.7
Calibration data
Each calibration of IQ sensors that are able to be calibrated causes an
entry to be made in the log book. Log book entries contain the following
information:
Calibration date
Calibration successful or not successful.
The detailed calibration data of the last calibrations are integrated in the
Calibration history of selected sensor
overview.
4.7.1 Calibration entries in the log book
The date and time of a calibration are entered in the log book. The
respective message text contains an indication as to whether a calibra
-
tion was successful or not. The values determined with calibration can
be viewed in the calibration history (see section 4.7.2).
Details on calibration are given in the operating manual for
the IQ sensor.
All calibration data are stored in the IQ sensor. In order to
view the calibration data of an IQ sensor, the IQ sensor
must be connected to the IQ S
ENSOR
N
ET
and must be
ready for operation.
1
Call up the measured value display with <M>.
2
Using <
>, highlight a sensor and confirm with <OK>.
The
Display/Options
menu opens.
3
Using <
>, highlight the menu item,
Log book of
selected sensor
, and confirm with <OK>. Calibration entries in
the log book are held in message code (ICxxxx and ECxxxx).
4
Select and open a calibration entry (ECxxxx or ICxxxx) with
<
> and <OK>.
5
Leave the
Calibration history of selected sensor
with
<
>.
Содержание IQ SENSOR NET System 2020 XT
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 8: ...Contents System 2020 XT CONNECT 0 6 ba75970e01 10 2012...
Страница 24: ...Overview System 2020 XT CONNECT 1 16 ba75970e01 10 2012...
Страница 28: ...Safety System 2020 XT CONNECT 2 4 ba75970e01 10 2012...
Страница 123: ...System 2020 XT CONNECT Settings setup 5 11 ba75970e01 10 2012 Fig 5 10 Edit list of sensors...
Страница 138: ...Settings setup System 2020 XT CONNECT 5 26 ba75970e01 10 2012...
Страница 152: ...Network connection System 2020 XT CONNECT 6 14 ba75970e01 10 2012...
Страница 178: ...PROFIBUS connection option System 2020 XT CONNECT 8 12 ba75970e01 10 2012...
Страница 180: ...Maintenance and cleaning System 2020 XT CONNECT 9 2 ba75970e01 10 2012...
Страница 194: ...What to do if System 2020 XT CONNECT 10 14 ba75970e01 10 2012...
Страница 204: ...Accessories and options System 2020 XT CONNECT 12 2 ba75970e01 10 2012...
Страница 211: ...Lists IQ SENSOR NET System 2020 XT CONNECT 13 7 ba75970e01 10 2012 W Weekly load diagram of selected sensor 4 19...
Страница 212: ...Lists IQ SENSOR NET System 2020 XT CONNECT 13 8 ba75970e01 10 2012...
Страница 215: ......