15.1 pH
Error message
OFL, UFL
The measured value is outside the measuring range.
Error message,
Error
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
IDS pH sensor:
– The measured value is outside
the measuring range of the
meter
– Use suitable IDS pH sensor
– Air bubble in front of the junction – Remove air bubble
(e.g. sway or stir the solution)
– Air in the junction
– Extract air or moisten junction
– Cable broken
– Replace the sensor
– Gel electrolyte dried out
– Replace the sensor
Cause
Remedy
IDS pH sensor:
– The values determined for zero
point and slope of the IDS pH
sensor are outside the allowed
limits.
– Recalibrate
– Junction contaminated
– Clean junction
– Sensor broken
– Replace the sensor
Buffer solutions:
– The used buffer solutions do not
agree with the set buffer set
– Set different buffer set
or
– Use different buffer solutions
– Buffer solutions too old
– Use only once.
Note the shelf life
– Buffer solutions depleted
– Change solutions