10.4 Cond measurement settings
10.4.1 Settings for IDS conductivity sensors
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the
<ENTER>
key. After completing the set-
tings, switch to the measured value display with
<M>
.
The possible settings are individually displayed for each sensor. Below the set-
ting menu is displayed for two IDS sensors (
TetraCon 925
,
LR 925/01
).
Default settings are printed in
bold
.
General setting
menu for
conductivity
Setting menu
TetraCon 925
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record of
the last calibration.
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the last calibration
records (max. 10)
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the calibration data stor-
age to the interface
Calibration
/
Calibration interval
1 ...
150
...
999 d
Calibration interval
for the IDS
conductivity sensor (in days).
The meter reminds you to cali-
brate regularly by the flashing
sensor symbol in the measuring
screen.
Reset
-
Resets all sensor settings to the
delivery condition (see section
10.7.1 R
Menu item
Possible
setting
Explanation
Type
Used measuring cell
Cal
Measuring cells whose cell con-
stants are determined through
calibration in the check- and cali-
bration standard 0.01 mol/l KCl
(1413 µS/cm @25 °C).
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant
is displayed in the status line.