inoLab
®
Multi 9310 IDS
Conductivity
ba75937e10
08/2021
57
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
8.1.2 Measuring the temperature
For reproducible conductivity measurements, it is essential to measure the
temperature of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in
the IDS sensor.
8.2
Temperature compensation
The calculation of the temperature compensation is based on the preset refer-
ence temperature, 20 °C or 25 °C. It appears on the display as
Tr20
or
Tr25
.
You can select one of the following temperature compensation methods:
Nonlinear temperature compensation
(
nLF
) according to EN 27 888
Linear temperature compensation (
Lin
)
with adjustable coefficient
No temperature compensation (off)
Application tips
Select the following temperature compensations given in the table according to
the respective test sample:
The reference temperature and temperature compensation are set
in the menu for the parameter, conductivity (see section 10.4.1
S
Test sample
Temperature compensation
Display
Natural water
(ground water, surface
water, drinking water)
nLF
according to EN 27 888
nLF
Ultrapure water
nLF
according to EN 27 888
nLF
Other aqueous solu-
tions
Lin
adjustable temperature coeffi-
cient 0.000 ... 10.000 %/K
Lin
Salinity (seawater)
Automatic
nLF
according to
IOT (International Oceano-
graphic Tables)
Sal
,
nLF
Содержание inoLab Multi 9310 IDS