pH value
inoLab
®
Multi 9310 IDS
36
ba75937e10
08/2021
5.2.6 Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
5.2.7 Calibration data
The calibration record of the last calibration is available in the menu
Calibration
/
Calibration record
. To open it in the measured value display, press the
<CAL__>
key.
The calibration records of the last 10 calibrations are available in the menu
Cali-
bration
/
Calibration data storage
. To open the
Calibration
menu in the
measured value display, press the
<ENTER>
key.
Calibration
Determined
values
Displayed calibration data
1-point
Asymmetry
Zero point =
Asymmetry
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point
Asymmetry
Slope
Zero point =
Asymmetry
Slope =
Slope
3-point to 5-
point
Asymmetry
Slope
Zero point =
Asymmetry
Slope =
Slope
The calibration line is calculated by
linear regression.
You can display the slope in the units, mV/pH or % (see section
10.1.1 S
The calibration record is automatically transmitted to the interface
after calibrating.