Ion concentration
93
7
Ion concentration
For the inoLab Level 3 measuring modules:
pH, pH/ION, pH/Cond, pH/Oxi and Multi.
7.1
Settings for ISE
We recommend checking the current settings and correcting
them if necessary before a measurement.
7.1.1
Displaying pH/ISE or measured voltage values
In pH/ISE measurements, inoLab Level 3 enables a rapid
changeover between the display of the pH/ISE value (
pH
)
and voltage (
U
).
7.1.2
Setting up the ISE display
Ion-selective electrodes are connected to the pH/U
connector of the measuring module.
The selection of the measured variable
ISE
or
pH
is made in
the main
Meas
menu. The active
pH
or
ISE
setting is marked
with
✔
.
Measured variable,
ISE
1
Activate the
Display
mode.
2
Switch to the display of the pH/ISE value with
m
.
3
Switch between the measured value display of the
pH/ISE value (
pH
) and voltage (
U
) with
46
.
1
Switch to the
Menu
mode.
2
Use
46
to select the main
Meas
menu.
3
Select the
ISE
menu item with
82
.
Содержание inoLab 740
Страница 18: ...Safety 18...
Страница 27: ...Commissioning 27...
Страница 32: ...Commissioning 32...
Страница 50: ...Basic principles of operation 50...
Страница 92: ...pH 92...
Страница 134: ...Ion concentration 134...
Страница 148: ...Dissolved oxygen D O 148...
Страница 174: ...BOD 174...
Страница 232: ...Data output 232...
Страница 236: ...Maintenance cleaning disposal 236...
Страница 270: ...Lists 270...
Страница 274: ...Lists 274...