Dissolved oxygen (D. O.)
137
8.1.3
Salinity correction
When measuring the concentration of test samples with a
salt content of more than 1 g/l, a salinity correction is
required. To do this, you have to enter the salinity equivalent
of the test sample (range 0.0 - 70.0) and to switch on the
salinity correction.
Entering the
salinity
7
Press
[ESC]
to close the air pressure display and
terminate the
Menu
mode.
1
Switch to the
Menu
mode.
2
Use
46
to select the main
Cal
menu.
3
Select the
Set Sal ...
menu item with
82
.
4
Confirm the selected menu item with
g
.
The input field for the salinity appears on the display.
5
Option:
Select a digit with
46
.
6
Increase or reduce the salinity value with
82
.
7
Use
[ESC]
to terminate the input.