Installation
DIQ/S 28X
26
ba77168d02
03/2017
If the DIQ/S 28X [-XX]/24V is operated at an ambient temperature below 47 °C
(117 °F), the operator has to make sure that the total power requirement P (con-
tinuous) is not more than 6.5 W.
If the DIQ/S 28X[-XX]/24V is operated at an ambient temperature above 47 °C
(117 °F), the allowed total power requirement P (continuous) is linearly reduced
to 4 W at 55 °C (131 °F).
Fig. 3-2
Degradation line of the 24V variant DIQ/S 28X[-XX]/24V
3.6
Connecting the sensor
The sensors are connected to a SENSORNET connector of the DIQ/S 28X.
General instal-
lation instructions
Lines must always be installed separately at a minimum distance of 20 cm from
other lines that carry a voltage greater than 60 V.
The free end of the sensor cable was stripped in the factory and all the wires are
fitted with wire end sleeves.
Tools
Phillips screwdriver
Small screwdriver.
NOTE
The sensor cable may only be connected to the SENSORNET connections. No
wire of the cable may be connected with an external electrical potential. Other-
wise, malfunctions could occur.
Power output [W]
Ambient temperature[°C]
52
54
56
48
50
-30
-20 .................. 0 ............ +46
Permissible ambient temperature -20 … +55 °C
54
56
0
4
4,5
5
5,5
6
6,5
7