Outputs
DIQ/S 28X
158
ba77168d02
03/2017
7.10.1 Switching on the maintenance condition
7.10.2 Switching off the maintenance condition
If the maintenance condition is activated for a sensor, the measured
values or status indicators of this sensor flash in the measured value
display.
1
Call up the measured value display with
<M>
.
2
Select the sensor you want to switch on the maintenance condition for
with
<
>
. The measured values of the sensor do not flash.
3
Open the
Display/Options
menu with
<OK>
.
4
Using
<
>
, highlight the menu item
Switch maintenance condi-
tion on/off
or
Maintenance Sensor S0x
and confirm with
<OK>
. A win-
dow that informs you about the maintenance condition opens up.
5
Confirm
Continue
with
<OK>
.
The selected sensor is in the maintenance condition. Linked outputs are
frozen.
6
Call up the measured value display with
<M>
.
The measured values of the sensor flash.
1
Call up the measured value display with
<M>
.
2
Select the sensor you want to switch off the maintenance condition for
with
<
>
. The measured values of the sensor flash.
3
Open the
Display/Options
menu with
<OK>
.
4
Using
<
>
, highlight the menu item
Switch maintenance condi-
tion on/off
or
Maintenance Sensor S0x
and confirm with
<OK>
. A win-
dow that informs you about the maintenance condition opens up.
5
Confirm
Continue
with
<OK>
.
The maintenance condition of the selected sensor is switched off.
Linked outputs are released.
6
Call up the measured value display with
<M>
.
The measured values of the sensor do not flash.