System 182
Safety instructions
2 - 3
ba75503e07
04/2012
Safe operation
If safe operation is no longer possible, the System 182 must be taken
out of operation and secured against inadvertent operation.
Safe operation is no longer possible if components:
have been damaged in transport
have been stored under adverse conditions for a lengthy period of
time
are visibly damaged
no longer operate as described in this manual.
If you are in any doubt, contact the supplier of your System 182.
Obligations of the
operator
The operator of the System 182 must ensure that the regulations and
guidelines listed below are followed when dealing with dangerous
substances:
EEC guidelines relating to safety at work
National laws relating to safety at work
Accident prevention regulations
Safety datasheets of the chemicals manufacturers.
Warning
A circuit (except for power supply connections and relay circuits),
that is connected to the System 182, must not feed any voltages
or currents that are not allowed. It has to be made sure that the
circuit at any time meets all requirements of a
Limited circuit
or
Limited Power
as well as of
SELV
(Safety Extra Low Voltage).
These include the following limiting value specifications:
AC voltage: max. 30 V effective / 42.4 V peak
DC voltage: max. 60 V
Current limit: max. 8 A
Power output limitation: max. 150 VA
Содержание DIQ/S 182-PR
Страница 6: ...Contents System 182 0 4 ba75503e07 04 2012 ...
Страница 10: ...Overview System 182 1 4 ba75503e07 04 2012 ...
Страница 14: ...Safety instructions System 182 2 4 ba75503e07 04 2012 ...
Страница 78: ...Operation System 182 4 18 ba75503e07 04 2012 ...
Страница 84: ...The SETTINGS menu System 182 5 6 ba75503e07 04 2012 ...
Страница 110: ...Relay outputs System 182 6 26 ba75503e07 04 2012 ...
Страница 122: ...PROFIBUS connection System 182 7 12 ba75503e07 04 2012 ...
Страница 124: ...Maintenance and cleaning System 182 8 2 ba75503e07 04 2012 ...
Страница 130: ...What to do if System 182 9 6 ba75503e07 04 2012 ...
Страница 142: ...Accessories and options System 182 11 2 ba75503e07 04 2012 ...
Страница 146: ...Appendix store separately if required System 182 13 2 ba75503e07 04 2012 ...