System 182
Technical data
10 - 3
ba75504e07
04/2012
Electrical data
DIQ/S 182-MOD/24V
(24 V AD/DC version)
Electrical connections
DIQ/S 182-MOD
(line power version)
Terminal strip inside the enclosure:
Electrical connections
DIQ/S 182-MOD/24V
(24 V AC/DC version)
Terminal strip inside the enclosure:
Relay
(3 x)
Supply
Nominal voltage:
24 V AC/DC ± 10 %
protective low voltage
SELV (Safety
Extra Low Voltage)
AC frequency:
50/60 Hz according to
DIN IEC 60038
Connection:
2 pin
Line cross-section of connections:
Europe: 1.5 ... 4.0 mm
2
USA:
AWG 14 ... 12
Fuse rating on the operator side: max. 16 A
Power consumption
max. approx. 12 W
SENSORNET 1
SENSORNET 2
RO
T
RO
T
RED
RED
SCHIRM
SCHIRM
SHIELD
SHIELD
GRÜN
GRÜN
GREEN
GREEN
X3
X6
X2
X5
X17
X9
X15
X13
X11
X1
X4
X16
X8
X14
X12
X10
100...
240V AC
NETZ/MAINS
AUXILIARY
R3
R2
R1
L1
N
240V AC
2A AC
£
£
240V AC
2A AC
£
£
240V AC
2A AC
£
£
HILFS-
SPANNUNG
AUXILIARY
VOLTAGE
SENSORNET 1
RO
T
RED
SCHIRM
SHIELD
GRÜN
GREEN
X3
X2
X17
X9
X15
X13
X11
X1
X16
X8
X14
X12
X10
24V AC DC
EINGANG
INPUT
POWER
AUXILIARY
R3
R2
R1
240V AC
2A AC
£
£
240V AC
2A AC
£
£
240V AC
2A AC
£
£
HILFS-
SPANNUNG
AUXILIARY
VOLTAGE
SENSORNET 2
RO
T
RED
SCHIRM
SHIELD
GRÜN
GREEN
X6
X5
X4
Output
Galvanically separated
Max. switching voltage
240 VAC or 24 VDC
Max. switching current
2 A (AC and DC)
Installation
requirements
Fuse rating on the operator side: max. 2 A
Relay functions
Programmable as:
– Opener or closer
– Limit monitor
– Monitoring of the warning and error
signals of the DIQ/S 182-MOD
– Proportional frequency output
– Proportional pulse width output
Содержание DIQ/S 182-MOD
Страница 6: ...Contents System 182 0 4 ba75504e07 04 2012...
Страница 10: ...Overview System 182 1 4 ba75504e07 04 2012...
Страница 14: ...Safety instructions System 182 2 4 ba75504e07 04 2012...
Страница 78: ...Operation System 182 4 18 ba75504e07 04 2012...
Страница 84: ...The SETTINGS menu System 182 5 6 ba75504e07 04 2012...
Страница 110: ...Relay outputs System 182 6 26 ba75504e07 04 2012...
Страница 126: ...Maintenance and cleaning System 182 8 2 ba75504e07 04 2012...
Страница 132: ...What to do if System 182 9 6 ba75504e07 04 2012...
Страница 144: ...Accessories and options System 182 11 2 ba75504e07 04 2012...
Страница 148: ...Appendix store separately if required System 182 13 2 ba75504e07 04 2012...
Страница 149: ......