MIQ/CR3; DIQ/CR3
Settings
4 - 39
ba64107e08
05/2006
4.8.2
Behavior in non-operative condition
An output is non-operative when no function is activated for the output.
An output becomes non-operative in case of
!
Power failure
(As soon as the supply voltage is sufficient again, the non-operative
condition of the outputs ends. The outputs function as specified by
the user again.)
!
Erasing a link to a sensor
!
Changing the
Measuring mode
sensor setting for a linked sensor
!
Changing the
Measuring range
sensor setting for a linked sensor
Note
Before editing sensor settings a note appears on the display to inform
you that links will be erased when you change the
Measuring mode
or
Measuring range
sensor setting.
Settings in the non-
active condition
Current output
Relay output
Current: 0 A
Relay: Open