AmmoLyt
®
Plus
700 IQ
Indexes
10 - 1
ba75651e03
08/2009
10
Indexes
10.1
Explanation of the messages
This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ S
ENSOR
N
ET
system for the AmmoLyt
®
Plus
700 IQ sensor.
Note
Information on
z
the contents and structure of the log book and
z
the structure of the message code
is given in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system
operating manual.
Note
The last three digits of the message code identify the source of the
message:
z
522 = AmmoLyt
®
Plus
700 IQ (armature / component class, ADA
adapters)
z
395 = AmmoLyt
®
Plus
700 IQ (ammonium / kalium sensor)
10.1.1 Error messages
Message code
Message text
EA1395
Meas. range exceeded or undercut
* Check process
* Select other meas. range
EA2522
Sensor temperature too high!
* Check process and application
EA3522
Sensor temperature too low!
* Check process and application
EAN395
Potassium measurement: range exceeded or undercut
* Check process
EIA522
Incorrect equipment
* for correct electrode equipment see operating manual
ES1522
Component hardware defective
* Contact WTW
Содержание AmmoLyt Plus 700 IQ
Страница 1: ...ba75651e03 08 2009 AmmoLyt Plus 700 IQ Operating manual IQ SENSOR NET Modular sensor for ammonium...
Страница 12: ...Safety instructions AmmoLyt Plus 700 IQ 2 4 ba75651e03 08 2009...
Страница 40: ...Matrix adjustment check and calibration AmmoLyt Plus 700 IQ 4 14 ba75651e03 08 2009...
Страница 42: ...Measuring AmmoLyt Plus 700 IQ 5 2 ba75651e03 08 2009...
Страница 48: ...Maintenance and electrode exchange AmmoLyt Plus 700 IQ 6 6 ba75651e03 08 2009...
Страница 62: ...Indexes AmmoLyt Plus 700 IQ 10 4 ba75651e03 08 2009...