AmmoLyt
®
Plus
700 IQ
Maintenance and electrode exchange
6 - 5
ba75651e03
08/2009
6.3
Exchanging the electrodes
Caution
The sensor can be damaged by dirt and moisture. Each time
before dismantling an electrode carefully clean the area around
the electrodes (section 6.2). Before mounting an electrode make
sure the area behind the sealing ring of the electrode and the
receptacle are dry and clean. The AmmoLyt
®
Plus
700 IQ may only
be submersed when the electrodes or original blind plugs are
mounted.
Use the special socket wrench provided to dismantle an electrode.
Electrodes are installed as described in section 3.4.1 E
Recognizing the
electrode type from
outside
When mounted, the electrodes can be recognized by the following
features:
Note
For the correct storage of the electrodes please follow the instructions
in section 3.3 N
Electrode
Hexagon
Front
surface
Other features
VARiON
®
Ref
black
black
– Thread at the
hexagon
VARiON
®
Plus
NH4
black
black
– Hexagon without
thread
VARiON
®
Plus
K
black
white
Содержание AmmoLyt Plus 700 IQ
Страница 1: ...ba75651e03 08 2009 AmmoLyt Plus 700 IQ Operating manual IQ SENSOR NET Modular sensor for ammonium...
Страница 12: ...Safety instructions AmmoLyt Plus 700 IQ 2 4 ba75651e03 08 2009...
Страница 40: ...Matrix adjustment check and calibration AmmoLyt Plus 700 IQ 4 14 ba75651e03 08 2009...
Страница 42: ...Measuring AmmoLyt Plus 700 IQ 5 2 ba75651e03 08 2009...
Страница 48: ...Maintenance and electrode exchange AmmoLyt Plus 700 IQ 6 6 ba75651e03 08 2009...
Страница 62: ...Indexes AmmoLyt Plus 700 IQ 10 4 ba75651e03 08 2009...