14 

World Precision Instruments

When pumping electrolytes, some indication of flow may be obtained by monitoring 
the resistance or the voltage of the micropipette. If, for example, the micropipette 
is filled with 3M KCl and the sample fluid is 0.1M KCl, a significant decrease in 
resistance of the micropipette occurs when pressure is applied. This happens 
because the resistance measured is determined primarily by the conductivity of 
the solution at and within the tip. As pressure is applied the 0.1M KCl surrounding 
the tip is replaced by the more highly conductive 3M KCl, decreasing the measured 
resistance. Likewise, if vacuum is applied, the resistance increases as the less 
conductive 0.1M KCl fills the tip.

NOTE

: The above technique will not work with exterior and interior solutions of the 

same concentration. However, a voltage change can be measured as a function of 
the pressure applied. Working with tip sizes of 1–2µm, we have seen a change in 
measured voltage of approximately 1mV for every 10 PSI of applied pressure.

Multibarrel Microinjection

For injection with multibarrel micropipette, the PolyFil multibarrel micropipette 
coupling kit can be purchased from World Precision Instruments. This multibarrel 
micropipette coupling kit allows easy and secure coupling of a multi-barrel 
micropipette to a pressure source. Kits include a five-port manifold which allows use 
of a single 

PV830

 to drive up to six micropipette barrels independently.

MAINTENANCE

The PicoPump has been designed to yield reliable performance. However particular 
laboratory conditions may require occasional replacement of the pressure and 
vacuum filters. Should this become necessary, return the instrument to the factory 
for filter replacement.

Cleaning

Do not use alcohol, aromatic hydrocarbons or chlorinated solvents for cleaning. 
They may adversely react with plastic materials used to manufacture the instrument.

The exterior of this instrument may be cleaned periodically to remove dust, 
grease and other contamination. There is no need to clean the inside. Use a soft 
cloth dampened with a mild solution of detergent and water. Do not use abrasive 
cleaners.

Содержание PicoPump PV830

Страница 1: ...PV830 Pneumatic PicoPump World Precision Instruments INSTRUCTION MANUAL Serial No _____________________ 042712 www wpiinc com...

Страница 2: ......

Страница 3: ...cking 2 INSTRUMENT DESCRIPTION 3 Setup 5 Vacuum Input 6 Pressure Input 6 Eject Pressure Port front panel 6 Vacuum Port 8 OPERATING INSTRUCTIONS 9 Front Panel 9 Rear Panel 10 Techniques In Microinjecti...

Страница 4: ...iv World Precision Instruments...

Страница 5: ...ally useful alternative to electroionophoresis since it does not mandate the use of charged ions Two different positive pressures may be applied one for ejection at high pressure and a second lower pr...

Страница 6: ...any parts are missing at 941 371 1003 or customerservice wpiinc com Returns Do not return any goods to WPI without obtaining prior approval RMA required and instructions from WPI s Returns Department...

Страница 7: ...position of the Vac Hold toggle switch This feature allows the pressure ejection channel to have a mild sucking action when pressure is not being applied A manual pneumatic switch also allows the vacu...

Страница 8: ...or Depress the plastic collar to remove tubing Fig 5 The PicoPump Pressure and Vacuum ports are located on the rear panel of the PV830 To install the hard tubing push it into the port until the tubing...

Страница 9: ...asily while metal tubing may require more effort For a diagram of the setup see Fig 6 Air or Nitrogen Supply 150 psi max Vacuum Pump or Line Pressure Input Vacuum Input Hold Eject Port Micropipette Ho...

Страница 10: ...o a nitrogen regulator on a N2 tank Recommended gases are dry air nitrogen or argon Never use corrosive gases If an air pressure line containing oil or water vapor is used an external filter is recomm...

Страница 11: ...hen using a MPH6S handle a firmer hold on the glass can be achieved by using two gaskets in the micropipette holder NOTE Capillary holders can be obtained from WPI which contain Ag AgCl half cells The...

Страница 12: ...ak when attaching or removing the tubing Before inserting the pipette place a gasket of the correct size in the cap Gasket Color Green Black Red White Pipette Diameter mm 1 0 1 2 1 5 1 65 WARNING DANG...

Страница 13: ...n dial yielding good pressure resolution The pressure regulating mechanism functions by continuously bleeding a small amount of gas This bleed rate increases as the difference between the input pressu...

Страница 14: ...electronic timer controls the time duration that the pressure solenoid stays open and the three commands serve only to start the timer Long pulses in the timed mode may be aborted by pressing the Stop...

Страница 15: ...helps you to correctly use the Hold pressure The backfilling flow rate is determined by the pressure of capillary action and the tip size Since the tip size is often determined by the requirement of t...

Страница 16: ...s cleaning the glass Small amounts of dust or grease can easily clog micron sized tips Cleaning with chromic acid solutions before pulling the electrode is commonly performed but care must be taken to...

Страница 17: ...ation The same technique may be used though if the drop is kept submerged under oil Droplets may seem to disappear after emergence from the tip Sometimes this is due to creepage of the aqueous fluid b...

Страница 18: ...approximately 1mV for every 10 PSI of applied pressure Multibarrel Microinjection For injection with multibarrel micropipette the PolyFil multibarrel micropipette coupling kit can be purchased from W...

Страница 19: ...e 1 Kit MPH6S for 1 5 mm pipette 5 ft tubing assembly 5430 20 PicoNozzle 1 Kit MPH6S for 2 0 mm pipette 5 ft tubing assembly 5430 ALL PicoNozzle 2 Kit MPH6S Micropipette Holder specify 1 0 1 2 1 5 or...

Страница 20: ...port will push the diaphragm and return the O ring to the seat To introduce a high pressure 1 Remove the 1 4 hard pressure tubing from the back of the pump 2 Use pieces of tubing with 1 4 OD to make a...

Страница 21: ...ng the knobs Over tightening the knobs will damage the regulators Vacuum Vacuum Input rear 0 30 0 in Hg 101kPa Vacuum Output 0 29 9 in Hg 101kPa Lowest Regulated Vacuum 3 in water 0 7kPa Regulator Acc...

Страница 22: ...hysical Specifications Power 95 130 VAC or 190 260 VAC switch selectable single phase 50 60 Hz 20VA Temperature 10 C 50 F to 40 C 104 F Dimensions 3 5 x 17 x 9 5 in 89 x 432 x 242 mm Mountable in stan...

Страница 23: ...L 215 m 267 m 1nL 100 m 1 24 m 100pL 46 m 57 m 10pL 21 5 m 27 m 1pL 10 m 12 4 m Fig 13 This graph compares the dimensions of a cube and a sphere with the same internal volume Volume of a Micropipette...

Страница 24: ...ion sensitive microelectrodes J Neuroscience Methods 8 231 247 1983 Deyhimi F and J A Coles Rapid silylation of a glass surface Choice of reagent and effect of experimental parameters on hydrophobici...

Страница 25: ...PV830 Pneumatic PicoPump World Precision Instruments 21...

Страница 26: ...22 World Precision Instruments DECLARATION OF CONFORMITY...

Страница 27: ...indirect special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused dir...

Страница 28: ...44 0 1462 424700 Fax 44 0 1462 424701 E mail wpiuk wpi europe com Germany Zossener Str 55 10961 Berlin Tel 030 6188845 Fax 030 6188670 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading...

Отзывы: