![WORLD OF MEDICINE A127 Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/world-of-medicine/a127/a127_instructions-for-use-manual_997764052.webp)
Technical Data
52 / 62
13
Technical Data
Type designation
A127
Product name
A127
Manufacturer information:
W.O.M. WORLD OF MEDICINE GmbH
Salzufer 8, 10587 Berlin
Software version
Can be determined by the authorized service technician
Mains voltage range [V]
100-240
Supply frequency range [Hz]
50 / 60
Fuse designation
2x T 3.15 A H 250 V
Power consumption
Current [A]
Power [VA]
Normal operation 100 V/60 Hz
1.15
115
Normal operation 240 V/50 Hz
0.59
141
Protection class (I, II)
I
Application part type (B, BF, CF)
BF
a
Defibrillator protection (yes/no)
No
Protection class (IP code)
IP41
Classification according to Medical Device Directive 93/42/
EEC, (I, IIa, IIb, III)
IIa
Conformity with the following standards: (in the currently
valid version)
IEC 60601-1/EN 60601-1
IEC 60601-1-2/EN 60601-1-2
Operating conditions
Temperature [°C] / [°F]
10 to 40 / 50 to 104
Relative air humidity [%]
30 to 75
Air pressure [kPa]
70 to 106
Max. altitude above sea level [m]
3000
Transport conditions
Temperature [°C] / [°F]
-25 to +70 / -13 to +158
Relative air humidity [%]
10 to 95
Air pressure [kPa]
70 to 106
Storage conditions
Temperature [°C] / [°F]
-25 to +70 / -13 to +158
Relative air humidity [%]
10 to 95
Air pressure [kPa]
70 to 106
Maximum sound level [dB]
< 80 dB(A) (with acoustic signals)
Suitable operating conditions (EMC)
Professional healthcare facility environment,
Suitability for use in environments with flammable anaes-
thetic mixtures (Category AP or APG)
This device is not designed for use with flammable anesthetic
agents (Category AP) or flammable anesthetic agents with oxid-
ants (Category APG)
Suitability for use in an oxygen rich environment [Yes /
No]
No
Mode of operation
Continuous operation
Approved sterilization / cleaning / disinfection method(s)
Wipe the surface of the device with a soft lint-free cloth
moistened with the surface disinfectant (for example Meliseptol
®
rapid). The concentration and the time the disinfectant must be
applied, depends on the information provided by the manufac-
turer of the disinfectant. Make sure moisture does not enter the
device.
Do not sterilize the device!
Max. time interval of safety check [a]
1
Max. output pressure [mmHg]
150
Max. irrigation rate [l/min]
1.5
Adjustable values/ranges
en
Содержание A127
Страница 1: ...A127 Pump for Arthroscopy A127 Instructions for Use en...
Страница 56: ...Appendix 56 62 15 Appendix 15 1 Test log Date Result Comment Signature en...
Страница 61: ......
Страница 62: ......