Device Setup
19 / 62
CAUTION!
Ventilation of the device
• Avoid device overheating.
• Ensure free air circulation especially to the bottom and rear of the device (rear
panel distance of at least 10 cm).
CAUTION!
Device setup
Device should be positioned outside of the sterile area in such a way that
• it can be easily disconnected,
• it is easy to use and switch off and on,
• it allows an easy monitoring of the display values, device functions, and access
to the control elements.
CAUTION!
Position of the user
To avoid a malfunction, the user must be positioned correctly
• within a display viewing angle of ±50° to operate the device,
• up to 2 m/6.5 ft from the device front for monitoring the actual values.
CAUTION!
To avoid malfunctions
The device A127 should not be used directly next to other devices as this could res-
ult in malfunctions. The device A127 was tested for compliance with IEC 60601-1-2
as a stand alone system. Therefore, do not stack other devices on the system or the
pump. If usage in the manner described above is nevertheless required, this system
and the other devices should be monitored to make sure they function properly.
Mains connection
WARNING!
Cable pulling
Always grasp the plug of the power cable when unplugging; never pull on the cable
itself.
CAUTION!
Connection to mains
Do not connect this device or system to a multiple socket-outlet or an extension
cord.