70 | UM-PP-EN-03
Fig. 35 WUS manual operation
Regardless of the current control status, it is always possible to query current measured values for
SPEED, INCLINE, TIME, DISTANCE, and PULSE. WOODWAY standard protocol commands are only
accepted when they are permitted in the current control mode. The only command which is also valid
in the program control status, external control status, and pulse control status is the RS-232 command
“STOP treadmill”.
Initial Status
In the initial status, SPEED and INCLINE are both set to zero. Current time and distance are maintained
(set to zero when restarting).
To return to initial status of WOODWAY standard protocol from all RUN modes:
•
Enter command “treadmill STOP”.
•
Repeat command once treadmill has stopped (STOP status, current speed = 0)
To return to initial status of WOODWAY standard protocol from STOP mode:
•
In STOP mode, press the STOP button.
•
Alternately, enter RS-
232 command “STOP treadmill” to return to initial status.
STOP Mode
The manual STOP mode can be reached from the RUN mode in one of the following ways:
•
Press the STOP button.
•
Enter the RS-
232 “STOP treadmill” command (interruption)
.
No speed commands are accepted in STOP mode. Current time, incline, and distance are maintained.
Restart is only possible once the treadmill has come to a standstill (press the [+] button or enter the
RS-
232 command “release treadmill” and then enter speed value).
To return to initial status from STOP mode, press the STOP button or enter RS-
232 command “STOP
treadmill”.
Содержание ST PPS 43 Med
Страница 22: ...16 UM PP EN 03 2 7 EC Declaration of Conformity Fig 1 EC Declaration of Conformity...
Страница 26: ...20 UM PP EN 03 3 5 Dimensions Fig 3 Device dimensions PPS Bari Mill 43 Fig 4 Device dimensions PPS Bari Mill 55...
Страница 27: ...21 UM PP EN 03 Fig 5 Device dimensions PPS Bari Mill 70 Fig 6 Device dimensions PPS Med 43...
Страница 114: ...108 UM PP EN 03 12 Maintenance Report DATE MAINTENANCE MEASURES FROM REMARKS...