62 | UM-PP-EN-03
WARNING
Danger Through Speeding-Up of the Running Surface!
If the drive motor is stopped (e.g. by pressing the STOP button,
emergency stop, or by power failure) when set at an incline, the
weight of the user (gravity) may cause the running surface to ac-
celerate.
►
Use special caution when stopping the drive motor when set at
an incline.
►
Users must be made aware of dangers before use.
Switch the device off via the main switch on the power supply console when it will not be used for a
long time.
Fig. 29 ON switch, PPS Ortho
1.
Emergency stop switch with magnet and pull-cord
2.
ON switch
ATTENTION
Do not move the running surface belt during the initialization phase
(approx. 3-4 seconds). The movement can be interpreted as a device
malfunction by the control electronics and the device will switch off.
►
Never step on the running surface during the initialization phase.
►
Do not leave the device until it switches back into STAND-BY
mode.
NEVER LEAVE THE TREADMILL UNATTENDED WHILE IT IS
SWITCHED ON.
2
1
Содержание ST PPS 43 Med
Страница 22: ...16 UM PP EN 03 2 7 EC Declaration of Conformity Fig 1 EC Declaration of Conformity...
Страница 26: ...20 UM PP EN 03 3 5 Dimensions Fig 3 Device dimensions PPS Bari Mill 43 Fig 4 Device dimensions PPS Bari Mill 55...
Страница 27: ...21 UM PP EN 03 Fig 5 Device dimensions PPS Bari Mill 70 Fig 6 Device dimensions PPS Med 43...
Страница 114: ...108 UM PP EN 03 12 Maintenance Report DATE MAINTENANCE MEASURES FROM REMARKS...