4.
The rotor must be loaded symmetrically.
5.
Do not use the centrifuge or rotor if it is not installed correctly or it is damaged.
6.
Do not use the centrifuge or rotor if it shows any sign of external mechanical
or electrical damages.
7.
The centrifuge must be used only for the specified application. It must not
be operated in a hazardous or flammable environment and no hazardous or
flammable content should be centrifuged.
8.
Do not fill tubes while they are in the rotor. Liquid spillage may harm the device.
9.
If liquid is spilled on the rotor or rotor chamber, the centrifuge must be cleaned
immediately.
10.
“Tube
Holders”
are a part of the Rotor. They should be placed in the rotor
whether 2,3,4, 6...12 tubes are being used.
11.
TROUBLESHOOTING
The centrifuge has a self-diagnostic function. If a problem occurs, an
error/warning code will be displayed on the screen and the operator can
determine the malfunction with the warning code below.
IMPORTANT NOTE:
1.
If the centrifuge is working slowly or is hot due to overdraw current, restart the
centrifuge and check it again.
2.
Do not use liquid with density higher than 1.2g/ml for full load operation.
12.
MAINTENANCE AND CLEANING
1.
The rotor and the outside of the centrifuge should be cleaned regularly with a
damp cloth.
2.
Ensure the centrifuge is not connected to power during cleaning.
3.
The brushless motor in the centrifuge does not require routine maintenance.
Any service required should be performed by Woodley Equipment Company
only. Repairs performed by unauthorised personnel may void the warranty.
4.
Always keep the centrifuge housing, rotor chamber, rotor and rotor
accessories clean. All parts should be wiped down periodically with a soft
cloth. For more thorough cleaning, use a neutral cleaning agent (Ph
between 6 and 8) applied with a soft cloth. Excessive amounts of liquid
should be avoided. Liquid should not come into contact with the motor.
5.
After cleaning, ensure that all parts are dried.
6.
Do not place the rotor into the cleaning solution
13.
WARRANTY STATEMENT
This product is warranted to be free from defects in material and workmanship
for a period of two (2) years from date of purchase. Your product will be duly
repaired upon prompt notification in compliance with the following conditions:
This warranty is valid only if the product is used for its intended purpose and
within the guidelines specified in this instruction manual. This warranty does
not cover damage caused by accident, neglect, misuse, improper service,
natural forces or other causes not arising from defects in original material or
workmanship. This warranty does not cover any incidental or consequential
damages, commercial loss or any other damages from the use of this product.
The warranty is invalidated by any non-factory modification, which will
immediately terminate all liabilities for the products or damages caused by its
use. The buyer and its customer shall be responsible for the product or use of
products as well as any supervision required for safety. If requested the
products must be returned in well packed and insured manner and all shipping
charges must be paid.
Some states do not allow limitation on the length of implied warranties or the
exclusion or limitation of incidental or consequential damages. This warranty
gives you specific legal rights. This warranty is given expressly in lieu of all other
warranties, expressed or implied.
The purchaser agrees that there is no warranty of merchantability or of fitness
for any intended purpose and that there are no other remedies or warranties,
expressed or implied, which extend beyond the description on the face of the
agreement. This warranty is only applicable to the original purchaser.
Products received without proper authorisation will not be entertained. All
items returned for service should be sent postage prepaid in the original
packaging or other suitable carton, padded to avoid damage. Woodley
Equipment Company will not be responsible for damage incurred by improper
packaging.
14.
PRODUCT DISPOSAL
In case the product is to be disposed of, the relevant legal regulations are to be
observed. The disposal of electrical devices is regulated within the European
Community by national regulations based on EU Directive 2012/19/EU on
waste electrical and electronic equipment (WEEE). According to these
regulations, any devices supplied after 13.06.05 in the business to business
sphere, to which this product is assigned, may no longer be disposed of in
municipal or domestic waste. Products are marked with the symbol
to indicate this.
Woodley Equipment Company Ltd
Old Station Park Buildings, St. John Street, Horwich, Bolton, BL6 7NY,
United Kingdom
Tel: +44 (0) 1204 669033
Fax: +44 (0) 1204 669034
www.woodleyequipment.com Email: [email protected]
User Manual
ERROR
PROBLEM
SOLUTION
No display
No mains power connection
Check power and properly plug in
power adaptor at both ends
Power failure
Check the mains power fuse
Improper connection
Connect adaptor properly
Lid open
Lid is
n’t
closed correctly
Close lid correctly
Error with lid closing and
opening mechanism
Contact Woodley Equipment Company
Err 55
Rotor not loaded symmetrically
Load the rotor symmetrically and
restart the centrifuge
Centrifuge lid
cannot be opened
Rotor is still spinning
Wait for the rotor to come to a stop
Power failure
Press emergency lid release after rotor
stops
Centrifuge shakes
during acceleration
& makes loud
running noises while
running
Rotor is not loaded
symmetrically
Load rotor symmetrically and
restart operation
Either a broken tube, damage to
the rotor or motor is the cause for
run noise
Replace broken tube. For damaged
rotor/motor, contact Woodley
Equipment Company
Rotor is damaged
Remove and change rotor
Display error
Loose connection of display
Contact Woodley Equipment Company
Err 1
Latch motor damaged, latch
jammed or any limit switch of latch
got damaged
Contact Woodley Equipment Company
Err 52
Rotor is stuck
Turn OFF the centrifuge, Check and fit
the rotor properly and turn ON the
centrifuge
Power tripping
Cable not fit properly
Remove cable and connect properly
Last run memory
not displayed
Turning ON centrifuge immediately
after turning it OFF
Maintain 3 seconds gap between
switching OFF and switching ON again
System freezes
Electronics error
Switch off centrifuge and then switch it
ON again. If the error still shows,
contact Woodley Equipment Company