Woodley Clinispin Micro II Скачать руководство пользователя страница 1

4.

 

The rotor must be loaded symmetrically. 

5.

 

Do not use the centrifuge or rotor if it is not installed correctly or it is damaged. 

6.

 

Do not use the centrifuge or rotor if it shows any sign of external mechanical 
or electrical damages. 

7.

 

The centrifuge must be used only for the specified application. It must not 
be operated in a hazardous or flammable environment and no hazardous or 
flammable content should be centrifuged. 

8.

 

Do not fill tubes while they are in the rotor. Liquid spillage may harm the device. 

9.

 

If liquid is spilled on the rotor or rotor chamber, the centrifuge must be cleaned 
immediately. 

10.

 

“Tube

 

Holders”

 are a part of the Rotor. They should be placed in the rotor 

whether 2,3,4, 6...12 tubes are being used. 

11.

 

TROUBLESHOOTING 

The centrifuge has a self-diagnostic function. If a problem occurs, an 
error/warning code will be displayed on the screen and the operator can 
determine the malfunction with the warning code below. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 
IMPORTANT NOTE:

 

1.

 

If the centrifuge is working slowly or is hot due to overdraw current, restart the 
centrifuge and check it again.

 

2.

 

Do not use liquid with density higher than 1.2g/ml for full load operation.

 

 

 

12.

 

MAINTENANCE AND CLEANING

 

1.

 

The rotor and the outside of the centrifuge should be cleaned regularly with a 
damp cloth.

 

2.

 

Ensure the centrifuge is not connected to power during cleaning.  

3.

 

The brushless motor in the centrifuge does not require routine maintenance. 
Any service required should be performed by Woodley Equipment Company 
only. Repairs performed by unauthorised personnel may void the warranty.

 

4.

 

Always  keep  the  centrifuge  housing,  rotor  chamber,  rotor  and  rotor 
accessories clean. All parts should be wiped down periodically with a soft 
cloth.  For  more  thorough  cleaning,  use  a  neutral  cleaning  agent  (Ph 
between  6  and  8)  applied  with  a  soft  cloth.  Excessive  amounts  of  liquid 
should be avoided. Liquid should not come into contact with the motor. 

5.

 

After cleaning, ensure that all parts are dried. 

6.

 

Do not place the rotor into the cleaning solution 

13.

 

WARRANTY STATEMENT 

This product is warranted to be free from defects in material and workmanship 
for a period of two (2) years from date of purchase. Your product will be duly 
repaired upon prompt notification in compliance with the following conditions: 
This warranty is valid only if the product is used for its intended purpose and 
within the guidelines specified in this instruction manual. This warranty does 
not cover damage caused by accident, neglect, misuse, improper service, 
natural forces or other causes not arising from defects in original material or 
workmanship. This warranty does not cover any incidental or consequential 
damages, commercial loss or any other damages from the use of this product. 
The warranty is invalidated by any non-factory modification, which will 
immediately terminate all liabilities for the products or damages caused by its 
use. The buyer and its customer shall be responsible for the product or use of 
products as well as any supervision required for safety. If requested the 
products must be returned in well packed and insured manner and all shipping 
charges must be paid. 
Some states do not allow limitation on the length of implied warranties or the 
exclusion or limitation of incidental or consequential damages. This warranty 
gives you specific legal rights. This warranty is given expressly in lieu of all other 
warranties, expressed or implied. 
The purchaser agrees that there is no warranty of merchantability or of fitness 
for any intended purpose and that there are no other remedies or warranties, 
expressed or implied, which extend beyond the description on the face of the 
agreement. This warranty is only applicable to the original purchaser. 
Products received without proper authorisation will not be entertained. All 
items returned for service should be sent postage prepaid in the original 
packaging or other suitable carton, padded to avoid damage. Woodley 
Equipment Company will not be responsible for damage incurred by improper 
packaging. 

14.

 

PRODUCT DISPOSAL 

In case the product is to be disposed of, the relevant legal regulations are to be 
observed. The disposal of electrical devices is regulated within the European 
Community by national regulations based on EU Directive 2012/19/EU on 
waste electrical and electronic equipment (WEEE). According to these 
regulations, any devices supplied after 13.06.05 in the business to business 
sphere, to which this product is assigned, may no longer be disposed of in 
municipal or domestic waste. Products are marked with the symbol  
to indicate this.  

 

Woodley Equipment Company Ltd 

Old Station Park Buildings, St. John Street, Horwich, Bolton, BL6 7NY,  
United Kingdom 
Tel: +44 (0) 1204 669033 

Fax: +44 (0) 1204 669034 

www.woodleyequipment.com  Email: [email protected] 

 

 

User Manual

 

 

 

 

 

 

 

 

 

 

 

 

ERROR

 

PROBLEM

 

SOLUTION

 

 

No display

 

No mains power connection

 

Check power and properly plug in 
power adaptor at both ends

 

Power failure

 

Check the mains power fuse 

 

Improper connection

 

Connect adaptor properly

 

 

Lid open

 

Lid is

n’t

 closed correctly

 

Close lid correctly

 

Error with lid closing and 
opening mechanism

 

Contact Woodley Equipment Company

 

Err 55

 

Rotor not loaded symmetrically

 

Load the rotor symmetrically and 
restart the centrifuge

 

Centrifuge lid 
cannot be opened

 

Rotor is still spinning

 

Wait for the rotor to come to a stop

 

Power failure

 

Press emergency lid release after rotor 
stops

 

 

Centrifuge shakes 
during acceleration 
& makes loud 
running noises while 
running

 

Rotor is not loaded 
symmetrically

 

Load rotor symmetrically and 
restart operation

 

Either a broken tube, damage to 
the rotor or motor is the cause for 
run noise

 

Replace broken tube. For damaged 
rotor/motor, contact Woodley 
Equipment Company

 

Rotor is damaged

 

Remove and change rotor

 

Display error

 

Loose connection of display

 

Contact Woodley Equipment Company

 

 

Err 1

 

Latch motor damaged, latch 
jammed or any limit switch of latch 
got damaged

 

 

Contact Woodley Equipment Company

 

Err 52

 

Rotor is stuck

 

Turn OFF the centrifuge, Check and fit 
the rotor properly and turn ON the 
centrifuge

 

Power tripping

 

Cable not fit properly

 

Remove cable and connect properly

 

Last run memory 
not displayed

 

Turning ON centrifuge immediately 
after turning it OFF

 

Maintain 3 seconds gap between 
switching OFF and switching ON again

 

 

System freezes

 

 

Electronics error

 

Switch off centrifuge and then switch it 
ON again. If the error still shows, 
contact Woodley Equipment Company 

 

 

Отзывы: