Settings
69
EN
10.6 Setting the O2-Parameters
(Probe not included in de-
livery)
•
Press and hold the Set key to access the
Settings mode, then use the arrow keys to
access parameter P 60.
•
Press Enter-Key to access P 6.1 mode.
•
P 6.1 – See Temperature
Calibration
Fig. 54: Temperature Calibration
Anzeige des letzten Temperaturkalibrierwertes:
Dieser Wert hilft zur Fehlerbeurteilung bei Feh-
leranzeige E21 im Display. E21 erscheint, wenn
die gemessene Temperatur um 10 ° C von der
kalibrierten Temperatur abweicht.
Review last temperature calibration value: This
value helps to judge cause when error E21 is
shown on the display. E21 appears if the meas-
ured temperature and the calibration tempera-
ture have 10 ° C difference.
P 6.2 Salinity Compensation
You can manually enter the Salinity Compensa-
tion.
•
Use salinity value measured by the conduc-
tivity probe.
•
Adjust the salinity value with the arrow keys.
(Range 0 bis 42.00 ppt).
•
Press Enter-key to confirm.
P 6.3 Height from Sea Level
•
Adjust the “hight from sea level” with the
arrow keys. (Range: 0 to 3550 m (100 m per
step)).
•
Press Enter-key to confirm.
P 6.4. Stabilization time
fast - slow
Default setting: FAS (fast )
•
SLOW: stabilization time 10 Minuten,
Advantage : Electrolyte consumption is low-
er and less maintenance work to do.
•
FAS (fast): reading stabilized in a few sec-
onds
Disadvantage : Electrolyte consumption is
much higher so the need to refill and doing
probe maintenance is more frequent.
10.7 Setting the temperature
unit
Default setting: °C
Options: °C or °F
•
Press and hold the Set key to access the
Settings mode, then use the arrow keys to