7.7
Reset of the temperature monitor for the mixed gas temperature (optional)
If the temperature of the mixed gas falls below the minimum temperature (see technical data) set at
the temperature monitor
(18)
the solenoid valve
(27)
in the mixed gas outlet closes.
Dependent on the build-in type of temperature monitor the reset must be done manually or occurs
automatically (see technical data).
Manual reset (standard):
If the mixed gas temperature exceeds the set minimum temperature for at least 3.5 °C proceed as
following:
Take the gas mixer out of operation.
Open the housing of the gas mixer.
Push the Push button of the temperature monitor to di-
rection specified right. Afterward release the push button.
Close the housing of the gas mixer.
Take the gas mixer into operation.
Automatical reset (special model):
If the mixed gas temperature exceeds the set minimum temperature for at least 3.5 °C an automat-
ically reset of the temperature monitor occurs.
The gas mixer automatically reassumes normal operation.