Page 6 of 12
Document No.: IM TMA227-15
Revision: A
TMM PUSH HANDLE
FIELD INSTALLATION AND USAGE INSTRUCTIONS
REMOVING AN OLD PUSH HANDLE ASSEMBLY
1. Using the pendant, adjust the device into the chair position
(Fig. 1)
.
2. Locate the bolt connecting the push handle to the back
frame. The position of this bolt will be slightly different on the
TMM4 and TMM5 Models
(Fig. 2 and Fig. 3).
NOTE: FOR THE TMM5 MODEL CHAIR, THE BOTTOM BOLT
OF THE SURGICAL BAR WILL NEED TO BE REMOVED AND
THE SURGICAL BAR WILL NEED TO BE MOVED OUT OF
THE WAY BEFORE CONTINUING TO THE NEXT STEP. THIS
WILL ALLOW ACCESS TO THE BOLT CONNECTING THE
PUSH HANDLE TO THE BACK FRAME (FIG. 3).
3. Using the 3/16” allen wrench and 1/2” socket wrench, remove
the bolt and fastener nut connecting the push handle to the
back frame
(Fig. 4)
.
4. Once the bolt and fastener nut are removed, locate the red
plunger knob and pull to release the old push handle from
the back frame
(Fig. 4)
.
5. Repeat this entire process on the opposite side of the device.
Fig. 1
Fig. 2
TMM4
Fig. 4
Fig. 3
TMM5
←
←
←