Press the menu soft-key
GPIB Address
to change
the current GPIB address of the sweep generator.
Enter a new address using the cursor control key or
the data entry keypad and the terminator key
Hz
ns
ADRS
The new GPIB address will now appear on the dis-
play. The entry must be between 1 and 30 to be rec-
ognized as a valid GPIB address.
Selecting
the Line
Terminator
Data is delimited on the GPIB by either the carriage
return (CR) ASCII character or both the carriage re-
turn and line feed (CR/LF) ASCII characters. Which
character is used depends upon the requirements of
the system controller. Most modern controllers can
use either CR or CR/LF, while many older control-
lers require one or the other. Consult the controller’s
manual for its particular requirements.
From the Configure GPIB Menu display, you can se-
lect which GPIB terminator to use by pressing the
menu soft-key
Line Term
. This menu soft-key tog-
gles the GPIB terminator between CR and CR/LF.
The current selection appears on the display.
GPIB SETUP AND
INSTALLATION
INTERCONNECTION
681XXA OM
2-9