26
7.3.2 Respiratory flow ranges
Operating mode
Tube diameter (mm)
Respiratory flow
range (l/min)
NIV
10
2–15
15
3–30
22
4–80
IV
10
2–15
15
3–30
22
4–60
FREE
10
2–15
15
3–30
22
4–80
7.3.3 Adjusting temperatures
This menu page is only displayed if the user has selected the
operating mode "FREE"� Notwithstanding the factory setting
for invasive and non-invasive ventilation, the user must set
the desired patient temperature and chamber temperature
manually here�
Note
In any operating mode (including IV and NIV), the temperature
can be adjusted to the needs of the patient (e�g� respiratory
gas too moist/too dry) if the button is pressed for about
3 seconds, so that the desired setting can be made� When
changing the temperature parameters, the device switches
automatically to "FREE" operating mode, even with factory
settings (IV or NIV)�
39.0
°C
36.0
°C