EN
14073818.10 05/2021 EN/DE/FR/ES
30
WIKA operating instructions model TR33
Output signal (4 ... 20 mA version)
Factory configuration
Measuring range
Measuring range 0 ... 150 °C [32 ... 302 °F]
Other measuring ranges are adjustable
Current signals for
error signalling
Configurable in accordance with NAMUR NE 043
downscale ≤ 3.6 mA
upscale ≥ 21.0 mA
Current value for
sensor short-circuit
Not configurable in accordance with NAMUR NE 043
downscale ≤ 3.6 mA
Communication
Info data
Tag no., description and user message can be stored
in transmitter
Configuration and
calibration data
Permanently stored
Configuration software
WIKAsoft-TT
→ Configuration software (multilingual) as a
download from www.wika.com
Voltage supply
Supply voltage U
B
DC 10 ... 30 V
Supply voltage input
Protected against reverse polarity
Permissible residual
ripple of supply voltage
10 % generated by U
B
< 3 % ripple of the output
current
Time response
Switch-on delay,
electrical
Max. 4 s (time before the first measured value)
Warm-up time
After approx. 4 minutes, the instrument will function
to the specifications (accuracy) given in the data
sheet.
12. Specifications
Содержание TR33
Страница 34: ...EN 14073818 10 05 2021 EN DE FR ES 34 WIKA operating instructions model TR33 ...
Страница 66: ...DE 14073818 10 05 2021 EN DE FR ES 66 WIKA Betriebsanleitung Typ TR33 ...
Страница 98: ...FR 14090084 10 05 2021 EN DE FR ES 98 WIKA mode d emploi type TR33 ...
Страница 130: ...14073818 10 05 2021 EN DE FR ES 130 WIKA operating instructions model TR33 ...
Страница 131: ...WIKA operating instructions model TR33 14073818 10 05 2021 EN DE FR ES 131 ...