WIKA operating instructions analog pressure transducer, model CPT6030
PN 0020508001A 03/2020 EN
EN
8.2 General Specification
Case
Orientation effects
Negligible – completely removable with re-zeroing
Dimensions
See technical drawings
Weight
~250 g (depending on range)
Instrument
Warm-up time
15 min.
Connections
Pressure connections
FSAE J514/JIC4 or Autoclave
®
F250C (for pressure ranges > 6000 psi)
10-32 UNF female port (for barometric range only)
Pressure port adapters
Standard: without
Option (only up to 6,000 psi): 1/8” female BSP fitting, 1/4” BSP fitting, 1/8” female NPT fitting, 1/4” male
NPT fitting, 6 mm tube fitting, 1/4” tube fitting and female 7/16-20 SAE fitting
Overpressure limit
2X proof, 3X burst; static pressure < 50 psig
Materials, wetted parts
Ranges ≤ 5 psi Silicon, 316 SS, silicon, glass filled resins, epoxy
Ranges > 5 to 1500 psi 316 SS
Ranges >1500 psi 316 SS, Fluorocarbon Rubber
Pressure media
Ranges
≤
5 psi – clean, dry, non-corrosive gases
Ranges > 5 psi – media compatible with the listed wetted parts.
Power
Voltage Range
15 to 28 VDC (24 VDC nominal)
Power consumption
4...20 mA depending on pressure input (23 mA, 0.65 W max.)
Permissible ambient conditions
Compensated temperature range
-20 to 75
o
C (-4 to 167
o
F)
Operating temperature range
-40 to 85
o
C (-40 to 185
o
F)
Storage temperature range
-40 to 85
o
C (-40 to 185
o
F)
Humidity
0 ... 95% r.h. (non -condensing)
Operating altitude
<3000 meters (10,000 feet)
Measure port internal volume
<1 cc
Reference port internal volume
~ 45 cc
Communication
Measuring rate
21 values/second
8.3 Certificates
Calibration
3)
Traceable to SI units through NIST or equivalent
3) Calibration in a vertical position.
Specifications