112
WIKA Betriebsanleitung Druckmessumformer, Typ IS-3
14243628.02 08/2018 EN/DE
DE
Alexander Wiegand SE & Co. KG
Control drawing type IS-3
Drawing no.
14137236.01
Page 2 of 4
Revision
created
checked
03
Date
31.05.2017
04.07.2017
02
Department IN-RD-M
PI-RD-CA
01
31.05.2017
By
Jens Wagner
Frank Ballweg
Non-incendive Parameters:
Vmax / Ui = 30 V, Imax / Ii = 100 mA at Tamb
≤
85 °C, Imax / Ii = 87 mA at Tamb > 85 °C, Pmax / Pi = 0.8 W
Ci = 16.5 nF (Flying Leads: + 0.2 nF/m), Li = 0 µH (Flying Leads: + 2 µH/m)
Non-Incendive Installation
Class I, Zone 2, Group IIC
Class I, Division 2, Groups A, B, C and D
Class II, III, Div. 2, Groups F, G
(Note 2)
Associated Apparatus
(Note 5), (Note 6), (Note 7)
Notes:
1. The non-incendive field wiring concept allows the interconnection of two devices with non-incendive parameters not specifically examined in combination as a system when:
Uo or Voc
≤
Vmax, Io or Isc
≤
Imax, Ca or Co
≥
Ci + Ccable, La or Lo
≥
Li + Lcable, Po
≤
Pi.
2. Dust-tight conduit seal must be used when installed in Class II and Class III environments.
3. Control equipment connected to the Associated Apparatus must not use or generate more than 250 Vrms or Vdc.
4. Installation should be in accordance with the Canadian Electrical Code (CEC) Part I for Canada or the National Electrical Code
®
(ANSI/NFPA70) Sections 504 and 505 for USA.
5. The configuration of Associated Apparatus must be under entity or non-incendive field wiring concept and - for the USA - FM Approved.
6. Associated Apparatus manufacturer`s installation drawing must be followed when installing this equipment.
7. Special Condition of use: Potential Electrostatic Charging Hazard - Parts of the enclosure may be constructed from plastic. To prevent the risk of electrostatic sparking the plastic
surface should be cleaned only with a damp cloth.
8. No revision to this drawing without prior approval by CSA and/or FM.
IS-3
Control Equipment
(Note 3)
Hazardous (Classified) Location
Non-Hazardous Location
Anlage 2: Control drawing FM, CSA
Содержание IS-3
Страница 58: ...58 WIKA operating instructions pressure transmitter model IS 3 14243628 02 10 2018 EN DE EN ...
Страница 110: ...110 WIKA Betriebsanleitung Druckmessumformer Typ IS 3 14243628 02 08 2018 EN DE DE Anlage 1 Konformitätserklärung ...
Страница 115: ...115 WIKA Betriebsanleitung Druckmessumformer Typ IS 3 DE 14243628 02 08 2018 EN DE ...