17
Operating instructions wire rope force transducer, model F9204
ADPR1X914116.01 04/2020 EN/DE
EN
7. Maintenance and cleaning / 8. Dismounting, return ..
Faults
Causes
Measures
Signal span varies
EMI sources in the environ-
ment; e.g., frequency
converter
Shield instrument; cable
shield; remove source of
interference
Signal span drops/too small
Mechanical overload
Consult the manufacturer
Setting behaviour of the wire
rope
Carry out readjustment
7. Maintenance and cleaning
For contact details see chapter 1 “General information” or the back page
of the operating instructions.
7.1 Maintenance
This instrument is maintenance-free.
Repairs must only be carried out by the manufacturer.
Only use original parts (see chapter 10 “Accessories”).
7.2 Cleaning
1. Prior to cleaning, disconnect the wire rope force transducers from the voltage supply
and dismount them.
2. Clean the wire rope force transducers with a cloth.
Electrical connections must not come into contact with moisture!
CAUTION!
Damage to the instrument
Improper cleaning may lead to damage to the instrument!
▶
Do not use any aggressive cleaning agents.
▶
Do not use any hard or pointed objects for cleaning.