6 Commissioning, operation
WIKA Operating Instruction, Model Pascal ET, ET-P, ET/IS, ET-P/IS 21
GB
6.1.2.3 Pressure module
The pressure module is installed only at factory location in accordance with the Pascal ET, ET-P,
ET/IS, ET-P/IS configuration required. It is possible to install one pressure modules, able to
support up to two transducers (2 internal, 2 external or a combination of both).
Transducer’s
selection with range and their resolutions are in the following table.
Internal transducers *
Range
Precision (% FS)
Accuracy (% FS)
Typical Resolution
Gauge
-
60…+60 mbar
0.1
0.15
0.001 mbar
-
500…+500 mbar
0.015
0.025
0.001 mbar
-
900…+1500 bar
0.015
0.025
0.01 mbar
0…7 bar
0.015
0.025
0,1 mbar
0…21 bar
0.015
0.025
0.1 mbar
0…50 bar
0.015
0.025
1 mbar
0…100 bar
0.015
0.025
1 mbar
Absolute
600…1300 mbar abs.
0.015
0.025
0.01 mbar
0…1500 mbar abs.
0.015
0.025
0.01 mbar
0…2500 mbar abs
0.015
0.025
0.01 mbar
0…5 bar abs.
0.015
0.025
0.1 mbar
0…7 bar abs.
0.015
0.025
0.1 mbar
0…21 bar abs.
0.015
0.025
0.1 mbar
0…81 bar abs.
0.015
0.025
1 mbar
0…100 bar abs.
0.015
0.025
1 mbar
External transducers *
Range
Precision (% FS)
Accuracy (% FS)
Typical Resolution
Gauge
-
60…+60 mbar
0.1
0.15
0.001 mbar
-
500…+500 mbar
0.015
0.025
0.001 mbar
-
900…+1500 bar
0.015
0.025
0.001 mbar
0…7 bar
0.015
0.025
0.1 mbar
0…21 bar
0.015
0.025
0.1 mbar
0…50 bar
0.015
0.025
1 mbar
0…100 bar
0.015
0.025
1 mbar
0…200 bar
0.015
0.025
10 mbar
0…400 bar
0.015
0.025
100 mbar
0…700 bar
0.025
0.05
100 mbar
0…1000 bar
0.025
0.05
100 mbar
Absolute
0…1500 mbar abs.
0.015
0.025
0.01 mbar
0…2500 mbar abs
0.015
0.025
0.01 mbar
0…5 bar abs.
0.015
0.025
0.1 mbar
0…7 bar abs.
0.015
0.025
0.1 mbar
0…21 bar abs.
0.015
0.025
0.1 mbar
0…81 bar abs.
0.015
0.025
1 mbar
0…100 bar abs.
0.015
0.025
1 mbar
* Other ranges available on request
6.1.2.4 HART module
The HART module allows communication with HART instruments (usually transmitters) permitting
to acquire digital measurements, instrument data and modifying settings. It can be inserted in the
front panel in place of OUT A or OUT B module.
It is a fully galvanically isolated board (like IN and OUT modules) from internal electronics of the
Pascal.
Содержание CPH8000
Страница 11: ...3 Specifications WIKA Operating Instruction Model Pascal ET ET P ET IS ET P IS 11 GB 3 Specifications 1...
Страница 12: ...3 Specifications 12 WIKA Operating Instruction Pascal ET ET P ET IS ET P IS GB...
Страница 13: ...3 Specifications WIKA Operating Instruction Model Pascal ET ET P ET IS ET P IS 13 GB...
Страница 14: ...3 Specifications 14 WIKA Operating Instruction Pascal ET ET P ET IS ET P IS GB...