![WHALETEQ MECG 2.0 Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/whaleteq/mecg-2-0/mecg-2-0_user-manual_967657029.webp)
MECG 2.0 Test System | User Manual
27
LoadWhaleteq_TXT
Waveform
file path
Waveform
file (*.txt)
path on your
local PC
Load the WhaleTeq
format waveform
(*.txt) file on local
PC.
LoadWhaleteq
_TXT
D:\WhaleteqM
ECG\MECG_227
ECG_.txt
StartPlay
N/A
Start outputting the
loaded ECG signal
data through the
MECG device.
StartPlay
StopPlay
N/A
Stop the current
signal outputting.
StopPlay
Continue
Seconds
Number of
seconds
before
continuing
next
command.
Pause the command
replay for specified
seconds then
continue.
Continue 200
Process
Executable
PathArg
The
executable
file path and
arguments
(optional).
Invoke an
executable to run.
Command replay
would continue next
command
immediately.
Process
NotePad.exe
result.txt
ProcessWait
Executable
PathArg
The
executable
file path and
arguments
(optional).
Invoke an
executable to run.
Command replay
would pause
until the running
process finished.
ProcessWait
NotePad.exe
result.txt
Below is a demo for command script:
Table 5: Demo for Command Script
Demo
Explain
CTS_CSE_Load PCTH009.CYC
Load CSE009 waveform
StartPlay
Start to output the waveform
Continue 144
Continue the behavior for 144 seconds
LoadLocal_Physionet
C:\Physionet\01.hea
Load Physionet
“
01
”
waveform from local PC
AcceptLeads II I
Lead mapping to Lead II, Lead I
StartPlay
Start to output the waveform
Continue 14
Continue the behavior for 144 seconds
StopPlay
Stop to output the waveform
7.7 Graph View
By default, the 12-lead ECG waveforms are all displayed in the main window.
To get more details of the waveforms, you can double-click one of the
waveform graphs to show the ECG Signal dialog.