MECG 2.0 Test System | User Manual
22
7.3 Load ECG Source from File
This function is set up to work with popular waveforms from the PhysioNet
website and other databases. As the software can work directly with the
website, the user does not need any knowledge about the PhysioNet site, file
formats and the like. However, note that there are many formats and options
available. The current MECG works with Format 16 and Format 212 with
common options.
Table 3: Load ECG Source from File
Use these buttons if the ECG files are
already on your PC.
For PhysioNet files, select the *.hea file
of interest. The *.dat file should be in the
same directory.
Use this section to automatically
download from the internet.
The downloaded file(s) will be stored in
“C:\Physionet”.
If the file has already been downloaded
before, the software will use the PC
version.
The import message log provides
feedback on what the software is doing
with the file. As many PhysioNet files
exceed ±5mV (due to noise, drift or large
physiological signals) and lead mapping
is not always clear, the user should
check these messages.
As there are many labels used for
waveforms in the PhysioNet website, the
MECG software will make a “best guess”
which output lead the waveform should be
mapped to. However, the user can adjust
these provided that selected leads are
exclusive.