71
AECG100 | User Manual
Figure 84: Self-Calibration Setup
3.
The AECG100
±
300 mV DC offset check is shown as the red square
in the figure below. The AECG100 connects DC in series with
electrode wire (in the case, RA).
Figure 85: AECG100 ±300 mV DC offset
4.
The user can verify the DC voltage by setting up the parameters of
the AECG100 AP to
“Square, 0 mV, 0.1 Hz and
DC Offset Setting to
300 mV”
.