34
AECG100 | User Manual
The PPG mode UI is for AECG100 main console unit with reflective SpO
2
module (PPG-2R-880/PPG-2R-940)/transmission SpO
2
module (PPG-
2TF-660)
Figure 33: PPG mode UI for reflective SpO
2
module and transmission SpO
2
module
(1)
BPM:
Adjust for different heart rates. The value is from 10 BPM to 300
BPM.
(2)
Output Waveform Parameter:
Adjust voltage and time parameters of
standard waveforms and pulse waveforms.
(3)
Output Waveform Selection:
Select to output different waveform
types (ex. PPG, Sine,
Triangle, Square…)
.
(4)
Noise Simulation:
To simulate noise with different frequency and
amplitude.
(5)
PPG Config:
Refer to 4.1 General (1) PPG Config.
(6)
Setting:
Refer to 4.1 General (2) Setting.
(7)
AC Offset:
Shift the signal upwards by the value of AC offset.
(8)
Respiration:
Add respiration modulation to the PPG function. Baseline
Modulation (BM), Pulse Amplitude Modulation (AM), and Frequency
Modulation (FM) are the available signals for breath modulation.
i.
Baseline Modulation (BM)
: Change in intrathoracic pressure when
breathing causes changes in venous blood return, which changes the
baseline of PPG
[1]
. Also called DC modulation.
ii.
Pulse Amplitude Modulation (AM)
: The heart axis deviates when
1
Addison, P.S., Watson, J.N., Mestek, M.L.
et al. Developing an algorithm for pulse oximetry derived respiratory rate (RR
oxi
): a
healthy volunteer study.
J Clin Monit Comput
26,
45
–
51 (2012).
https://doi.org/10.1007/s10877-011-9332-y
9
1
5
6
2
3
4
7
8
10