background image

14

EN

7.4.  Resolution 

The number of pixels used by the CMOS line is defined by the resolution. The higher the resolution set, the 
smaller the detectable movements of an object and the slower the output rate of the Sensor becomes when 
there is a change in the measured values.

Resolution

Adjust the display device

 High

 Medium

 Low 

3

 Back

7

 Run

High:  

High resolution, low output rate

Medium:    Medium resolution, medium output rate.
Low:  

Low resolution, high output rate.

7.5.  Filter 

The filter (filter size) is the number of measured values over which the Sensor takes an average. The larger the 
filter selected, the slower the response time of the Sensor becomes when there is a change in the measured 
values. A larger filter improves the reproducibility of the Sensor.

Filter

Number of values for averaging

 1 

 2 

 5 

 10 

 20 

 50 

 100 

3

 Back

7

 Run

If 1 is selected, each measured value is output directly without averaging. When-
ever a value greater than 1 is selected, the Sensor takes an average over the 
selected number of x measured values.

7.6.  Exposure 

The Sensor adjusts its exposure time or light pulse duration automatically up to a maximum value on the object 
to be detected. In the default DCM (Default Capture Mode), the Sensor has a fixed maximum possible exposure 
time. In the case of black or shiny objects, for example, it can be helpful to lengthen this time. It can be helpful to 
reduce the exposure time when the Sensor is aimed at the light source. In the case of problematic applications, 
it can be influenced by different modes or manually.
 

Exposure

Adjusting Exposure Times

 SCM 

 LCM 

 DCM 

 CCM 

3

 Back

7

 Run

SCM (Short Capture Mode):     The Sensor exposure time can be shortened on 

dark or glossy objects (e.g. black varnish) in order 
to reduce a fall in the measuring rate.

LCM (Long Capture Mode):     The Sensor exposure time can be lengthened on 

dark or glossy objects (e.g. black varnish) in order 
to achieve a more accurate measurement.

DCM (Default Capture Mode):   Default exposure time for standard applications.
CCM (Custom Capture Mode):  The Sensor exposure time can be adjusted manu-

ally between 100…2000 µs.

Содержание OCP P0150P Series

Страница 1: ...OCPxxxP0150P High Performance Distance Sensor Operating instructions Available as PDF only Status 27 01 2016 www wenglor com EN...

Страница 2: ...4 4 Complementary Products 8 5 Mounting Instructions 8 6 Initial Operation 9 6 1 Initial Operation 9 6 2 Default Settings 9 7 Functional Overview OLED Display 10 7 1 Run 12 7 2 Display 12 7 2 1 Displ...

Страница 3: ...ed configuration 24 10 1 Invoking the Management Interface 24 10 2 Page layout 25 10 3 Device general 26 10 4 Device Settings 27 10 5 Measured Value Settings 28 10 6 Device Test 29 11 Maintenance Inst...

Страница 4: ...s Furthermore Teach In filter functions for adjusting a switching output and an error output are available The measuring range can be selected individually within the working range 2 Safety Precaution...

Страница 5: ...ght 10000 Lux Beam Divergence 2 mrad Light Spot Diameter 3 6 0 9 mm Port Type 100BASE TX PoE Class 1 Temperature Range 25 50 C Reverse polarity Protection Yes Interface PROFINET Protection Class III A...

Страница 6: ...80 240 300 360 420 480 540 600 660 1000 70 0 Working Range mm Reproducibility m OCP662P0150P Typical Linearity Error Curves within the Working Range 30 35 40 45 50 55 60 65 70 75 80 100 50 0 Working R...

Страница 7: ...L Black Brown Red Orange Yellow Green Blue Violet Grey White Pink Green Yellow Supply Voltage Supply Voltage 0 V Supply Voltage AC Voltage Switching Output NO Switching Output NC Contamination Error O...

Страница 8: ...nction may be activated via Engineering Tool L A Green Links present Green flashing Communication 4 4 Complementary Products wenglor offers Connection Technology for field wiring Suiting Mounting Tech...

Страница 9: ...tage supply When the voltage supply is present the display on the Sensor starts Install the associated device specific electronic description file e g for PROFINET the GSDML file in the hardware manag...

Страница 10: ...twork Mode Rotate Intensity Display Offset Filter Resolution 1 10 100 2 20 5 50 SCM LCM DCM CCM Laser Illumination On Off E A Test T for Teach In Z for reset Value to 0 Value in m Run Mode Value in m...

Страница 11: ...tion menu by pressing any button Note If no setting is made in the configuration menu for a period of 30 s the Sensor returns automatically to the display view Pressing the button again returns the Se...

Страница 12: ...y 180 The rotation is canceled by pressing this button again Intensity Set the display intensity see chapter 7 2 2 7 2 1 Display Mode The measured value in mm and the PROFINET device name always appea...

Страница 13: ...value is adopted as the default offset value by pressing the T The offset value is set to 0 by pressing the Z button Change Change the value of the offset By pressing the button or the button the val...

Страница 14: ...lected the Sensor takes an average over the selected number of x measured values 7 6 Exposure The Sensor adjusts its exposure time or light pulse duration automatically up to a maximum value on the ob...

Страница 15: ...F Switch transmitted light off the Sensor no longer supplies mea sured values 7 9 I O Test This function manually changes the output of the Sensor As a result it possible to test whether the further p...

Страница 16: ...ured value is reduced as a result Current measured value 7 Warm up procedure The sensor is currently in the warm up phase and the quality of the measured value does not yet comply with the speci fied...

Страница 17: ...ction 7 12 Info In the menu item Info the following information about the Sensor is displayed Info Order number Software version Serial number 7 13 Reset All Sensor settings with the exception of the...

Страница 18: ...iate disabling of operation if activate Password is set to on If the password functionality is activated the password must be entered before each operation of the Sensor After correctly entering the p...

Страница 19: ...1 Byte 0 Min 1 Normal 2 Max 3 Power saver mode 4 Screensaver Display mode BitArea 0 0 3 0 Process 0 3 Yes 304 1 Byte 0 Process 1 Analysis 2 Text 3 Network Language BitArea 0 0 3 1 English 0 4 Yes 305...

Страница 20: ...8 257 0 Yes Resolution 0 2 Unsigned8 258 0 Yes Exposure Mode 0 3 Unsigned8 259 2 Yes Exposure Time 100 2000 Unsigned16 260 1000 Yes Laser 0 1 Bit 264 0 Yes Module Status Module ID 2 Submodule 1 Slot 2...

Страница 21: ...1 1 filter size 2 2 filter size 5 3 filter size 10 4 filter size 20 5 filter size 50 6 filter size 100 Parameter Name Value Range Data Type Index Module Laser ON OFF Module ID 4 Submodule 1 Slot 4 For...

Страница 22: ...e ID 6 Submodule 1 Slot 6 Format Cyclical Data Output Value Range Data Type Unit Description 1 5 Unsigned8 0x00 Start command 0x01 Learn offset value 0x03 Reset offset value 0x04 Apply offset 0x05 Res...

Страница 23: ...ned8 0 Resolution high 1 Resolution medium 2 Resolution low Parameter Name Value Range Data Type Index Module Exposure Mode Module ID 8 Submodule 1 Slot 8 Format Cyclical Data Input Input Output Value...

Страница 24: ...structions the default values are always assumed WARNING It must be ensured that supply power is not interrupted while making changes to network settings Furthermore supply power must maintained for a...

Страница 25: ...erman French Span ish or Italian via the language selection 2 Display On each page the current display is represented exactly like on the device itself 3 Category selection The web based settings are...

Страница 26: ...the overview page Device general is displayed Device Name A unique device name can be freely assigned to each device via the PROFINET control The device name is also displayed in the first line of th...

Страница 27: ...tings are reset to the factory settings See chapter 6 2 Default Settings on page 9 The settings are applied after interruption of the power supply The Sensor settings remain unchanged Display Settings...

Страница 28: ...n Reset the offset function is reset and the actual distance is displayed For functional description of offset see chapter 7 3 Offset on page 13 Resolution For functional description of resolution see...

Страница 29: ...below the display window The test can also be terminated prematurely by clicking on Switch off 11 Maintenance Instructions This wenglor Sensor is maintenance free It is recommended to clean the lens a...

Отзывы: