6.2.5 Filter
• The filter represents the number of values used by the sensor to generate a mean value for pressure mea-
surement.
• The higher the filter number, the longer the sensor’s step response time T90 when the measured values
change.
Filter
Step Response Time T90
0
8 ms
1
16 ms
2
32 ms
3
66 ms
4
135 ms
5
270 ms
6
500 ms
7
1,0 s
8
2,1 s
9
4,3 s
10
8,4 s