Wellue PO6 Скачать руководство пользователя страница 1

Pulse Oximeter

Instruction for Use

1. Introduction

1.1

Intended use

This Pulse Oximeter is intended to be used for measuring,

displaying and storing of pulse oxygen saturation (SpO

2

),

pulse rate of adults in home or healthcare facilities

environment.

1.2

Contraindications

No contraindications.

1.3

Warnings and Cautions

DO NOT squeeze the sensor part or apply excessive
force on it.

Do not use this device during MRI examination.

Do not use this device with a defibrillator.

Do not store the device in the following locations:

locations in which the device is exposed to direct

sunlight, lint,dust,high temperatures or levels of

moisture, or heavy contamination; locations near to

sources of water or fire; or locations that are subject to

strong electromagnetic influences.

Do not use the device in a combustible environment

(i.e., oxygen-enriched environment).

Never submerge the device in water or other liquids.

Do not clean the device with acetone or other volatile

solutions.

Do not drop this device or subject it to strong impact.

The device and accessories are provided non-sterile.

Do not place this device in pressure vessels or gas

sterilization device.

Do not dismantle the device, as this could cause

damage or malfunctions or impede the operation of the

device.

Consult your doctor immediately if you experience

symptoms that could indicate acute disease.

Do not self-diagnose or self-medicate on the basis of

this device without consulting your doctor. In particular,

do not start taking any new medication or change the

type and/or dosage of any existing medication without

prior approval.

Use only cables, sensors and other accessories

specified in this manual.

Prolonged continuous monitoring may increase the risk

of undesirable changes in skin characteristics, such as

irritation, reddening, blistering or burns.

Do not open the device cover without authorization.

The cover should only be opened by a qualified service

personnel.

The biocompatibility testing has been performed on the

materials in contact with the person in accordance with

ISO10993.

Do not place the SpO

2

probe on a finger with edema or

fragile tissue.

Check the SpO

2

sensor and cable before use. Do not

use a damaged SpO

2

sensor.

Check the SpO

2

sensor application site every 6-8 hours

to determine the positioning of the sensor and the

circulation and skin sensitivity of the patient. Patient

sensitivity varies depending on medical status or skin

condition. For patients with poor peripheral blood

circulation or sensitive skin, inspect the sensor site

more frequently.

The functional tester cannot be used to assess the

accuracy of the SpO

2

sensor or a device.

The device has no alarm system.

Continuous use for a long time may cause allergies,

redness, blistering or burns. Check the wearing

position every 6-8 hours.

The local laws and regulations should be followed

when disposing of the device and accessories.

Do not maintain the device while it is charging.

Please keep the cable away from children. It can cause

strangulation.

Keep the device out of reach of pets, pests and

children.

The PULSE OXIMETER EQUIPMENT is calibrated to

display FUNCTIONAL OXYGEN SATURATION.

1.4

Guide to Symbols

Symbol

Description

Type BF-Applied Part
Manufacturer

Date of manufacture

Authorized Representative in the

European Community
Follow Instructions for Use.
MRI unsafe. Presents hazards in all MR

environments as device contains strongly

ferromagnetic materials.

IP22

Protected against spraying water and

against access to hazardous parts with a

tool, per IEC60529.

SN

Serial number

No alarm system

Temperature limitation

Humidity limitation

Atmospheric pressure limitation
Indicate separate collection for electrical

and electronic equipment (WEEE).

1.5

Unpacking

Device

User Manual

Charging Cable

2

Overview

Name:Pulse Oximeter

Model: PO6

PO6A

Model and Configuration see the table below.

3

Using the Device

3.1

Charging

Use the USB cable to charge the product. Connect

the USB cable to a USB charger or to the PC.

There will be a battery logo flash on the device

when charging.When the battery is low,the display

screen will display a low power prompt.

A fully charge will need 2-3 hours. After fully

charged, the device will power off automatically.

Note:

The device cannot be used during charging,

and if choosing a third party charging adaptor

(Class II), select one that complies with IEC60950

or IEC60601-1.

3.2

POWER ON/OFF

POWER ON:

Wear the device, it will turn on automatically.

POWER OFF:

The device turns off automatically in a moment

after you take off the sensor.

3.3

Typical steps

1)

START.

Charge the battery. Wear the device

to power on.

2)

STOP.

Take off the device, the recording will

be over after the countdown.

3)

DATA Uploading.

Run your phone to download data.

3.4

Start working

1)

Wear the device on index finger. Try to

move the device along the forefinger to find

out a best fit. Avoid being loose. Loose

wearing causes inaccurate measure.

2)

Device will turn on automatically. After a few

seconds, the device will begin to work.

Notice:

If the working time is less than 2 minute, the

data will not be saved.

Please avoid excessive motion.

Please avoid strong ambient light condition.

SpO

2

measurement principle:

The Pulse Oxiemter is a lightweight, portable

health oximeter for use in the home or in

healthcare

facilities.

SpO2

measurement

technology is based on developed photoelectron

method, the circuit design and calculation software

was developed by Shenzhen Viatom Technology

Co., Ltd. The SpO2 sensor receives the optical

signal from the red light and infra-red light through

the finger. Insert the finger into the oximeter, there

No.

Model

Appearance

color

1

PO6

Blue

2

PO6A

Gray