2.7 Emergency Shutoff
Figure 11: Packing Adjustment Nut
1.
If sample begins to free flow from the sample outlet, turn OFF the hydraulic or pneumatic supply to the motor, and then
close emergency shutoff valve E.
DO NOT operate the pump after emergency shutoff valve E is closed. Operating the pump while emergency shutoff valve E is
closed may damage the pump.
Free flow from the sample outlet indicates that the external adjustable relief has not been set correctly or that the external
adjustable relief has failed.
2.
If leaking becomes evident elsewhere on the inFlow™ ACE, turn OFF the hydraulic or pneumatic supply to the motor,
and then tighten the packing adjustment nut to stop the leak until the unit can be removed for service.