Appendix
Specifications
Physical specifications
Protection classifications, all device configurations
Characteristic
Specification
Electrical rating
100 – 240 V AC, 50 – 60 Hz, 0.8A – 0.5 A
Duty cycle
Continuous operation
Type of protection against electric shock
Class I internally powered
Degree of protection against electric shock, for parts
applied to patients
Type BF defibrillator proof
IEC EN 60601-1, 2nd and 3rd Editions
Recovery time following defibrillator discharge
Less than or equal to 15 seconds
Flammable anesthetics
WARNING
Not suitable for use
with flammable anesthetics.
Degree of protection provided by the enclosure with
respect to harmful ingress of liquids
IPX2 Protected against vertically falling water drops
when enclosure tilted up to 15°
Height
10.1 in. (25.7 cm )
Width
9.3 in. (23.6 cm )
Depth
4.9 in. (12.4 cm )
Weight (including battery)
3.8 lb (1.7 kg )
Graphical display resolution
Dimensional outline
6.5 in. (W) x 4.1 in. (H) x 0.13 in. (D) (164.9 mm [H] x
103.8 mm [W] x 3.40 mm [D] )