Remove the power supply from the mobile stand
1. Power down the device.
2. Disconnect the power cable from the device and disconnect the power cord from the wall.
3. With one hand holding the power supply, remove the 2 Phillips screws that attach the power
supply to the bracket.
NOTE
Set the screws aside for reassembly.
4. Remove the power supply unit.
Recycle or dispose of the power supply unit in an environmentally safe manner consistent with
local regulations.
Attach the power supply to the stand
Kit item:
•
107211, SRV KIT, POWER SUPPLY
1. Ensure that the side with the screw holes is facing upward, and then insert the power supply
socket into the support frame.
2. Holding the power supply unit snugly to the frame, attach the power supply to the bracket
using the 2 Phillips screws.
Service manual
Disassembly and repair 59
Содержание Spot Vital Signs 4400
Страница 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Страница 8: ...4 Symbols and definitions...
Страница 34: ...30 Power up sequence...
Страница 68: ...64 Field replaceable units...
Страница 90: ...Material no Item 4400REPRC WEB TRN 4400 Repair re certification web training 86 Options and upgrades...