4. Grasp the device securely and remove the device from the mount.
Reassembly notes
NOTE
Remounting the device is the reverse of removal.
Remove the device from the 77794-M4400 or
77794-2M4400 Integrated Wall Panel
1. Disconnect the power cord from the wall.
2. Disconnect the power supply and communication cables from the device.
3. Detach the device from tilt tensioner of the wall board.
a. Loosen the screw that secures the device to the tilt tensioner.
b. Firmly grasp the device with both hands and pull up to remove the device from the
bottom and top mounting hooks of the wall-mount bracket.
38 Disassembly and repair