Keep if + or -
Specify the range that the program uses as criteria
to accept or reject readings and establish the
baseline reading.
The configured settings are displayed in the
Summary
section.
3. Optionally, configure other advanced settings.
4. When you are finished configuring advanced settings, touch
Exit
to return to the
Home
tab.
Service tab
Perform general device service
You can calibrate the touchscreen and check the functionality of NIBP accuracy on the device
without use of the Welch Allyn Service Tool.
1. In Advanced settings, touch the
Service
>
General
tabs.
2. To verify the NIBP pressure valve accuracy, touch
Close Valve
.
The pressure displays in the NIBP Pressure Verification field.
To stop the accuracy test, touch
Open Valve
or navigate away from the screen.
3. To calibrate your touchscreen, touch
Start
in the right pane and follow the onscreen prompts.
The
Home
tab appears when calibration is complete.
Save service logs to a USB flash drive
This procedure enables you to save the last 14 active log files to a USB flash drive.
1. In Advanced settings, touch the
Service
>
Log file
tabs.
2. Connect a flash drive to the USB port.
3. Touch
Save to USB
.
Copies of the service log files are saved to the drive. You may remove the flash drive from the
device.
Restore factory default settings
NOTE
This process does not restore the device to a custom configuration
provided at delivery. Instead, it deletes all custom configuration data and restores
factory default settings.
1. In Advanced settings, touch the
Service
>
System
tabs.
2. Touch
Reset
.
3. Click
OK
to confirm.
The device reboots and the default factory settings are restored.
26 Advanced settings
Содержание Spot Vital Signs 4400
Страница 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Страница 8: ...4 Symbols and definitions...
Страница 34: ...30 Power up sequence...
Страница 68: ...64 Field replaceable units...
Страница 90: ...Material no Item 4400REPRC WEB TRN 4400 Repair re certification web training 86 Options and upgrades...
Страница 127: ...Interconnect diagram Service manual Appendix 123...
Страница 128: ...124 Appendix...
Страница 132: ......