Strabismus Threshold Value(age): Range 0 to +20 degrees in increments of 1
degree.
To modify current screening criteria
1. Perform an export and insert the USB flash drive into your computer.
2. Go to the USB drive folder and open the Spot folder with the most recent
timestamp. Select the import subfolder.
3. Copy the SpotCriteria.csv file to the top level on the USB drive. The file cannot be
within a folder. Once copied, open the file.
4. Make any desired changes using a spreadsheet or notepad program.
5. Save file as a Comma Separated Value (.csv) file. Do not change the file name. Eject
the USB drive from your computer and insert into the vision screener.
6. Touch
Tools
, and then touch
Import/Export
.
7. Touch
Import
to import the updated criteria file.
8. To verify the file was imported successfully, touch
Tools
, and then touch
Criteria
.
Review the new settings.
9. To revert to the default criteria that was shipped with the device, touch
Restore
twice.
The custom criteria is deleted from the vision screener and the default criteria is
restored.
About licensing
To use the screening functionality of the vision screener, the device needs to be
licensed. If your device came new from the factory with this Directions for use, your
license is already installed and active. If you encounter an issue with the license, contact
Welch Allyn Support at [email protected]
To enable/disable security on the vision screener
If you want to prevent unauthorized access, you can enable security on the vision
screener by adding a four-digit security PIN code
CAUTION
If you create a security PIN, you cannot disable security on the
device without the PIN. If you lose or forget the PIN, contact customer
service. The device must be restored to factory settings. Any data on the
device will be lost.
1. Touch
Tools
, then touch
Security
.
2. Check Use security PIN code.
3. Enter and then re-enter the security PIN code.
4. To disable security, touch
Tools
, then touch
Security
.
Uncheck Use security Pin code.
5. Enter the security PIN code.
6. To return to the previous menu, touch the
X
in the upper-left corner of the screen.
Directions for use
Advanced settings on the vision screener 35
Содержание Spot Vision Screener VS100
Страница 1: ...Welch Allyn Spot Vision Screener Model VS100 Directions for use Software version 3 1 XX ...
Страница 16: ...12 Controls Display Window and Connections Welch Allyn Spot Vision Screener Model VS100 ...
Страница 20: ...16 Setting up the vision screener Welch Allyn Spot Vision Screener Model VS100 ...
Страница 28: ...24 Screening subjects Welch Allyn Spot Vision Screener Model VS100 ...
Страница 48: ...44 Troubleshooting Welch Allyn Spot Vision Screener Model VS100 ...
Страница 60: ...56 Service policy Welch Allyn Spot Vision Screener Model VS100 ...
Страница 62: ...58 Warranty Welch Allyn Spot Vision Screener Model VS100 ...
Страница 64: ...60 Configuration options for the VS100 vision screener Welch Allyn Spot Vision Screener Model VS100 ...
Страница 68: ...Material No 728900 ...