Page 23
SP-2 User Guide
Art. No. 71022-0000 Welch Allyn Schiller ©2000
E
n
g
l
i
s
h
Spiro Settings (Printout)
In the printout settings screen the information given on a printout is defined.
Select between:
Format
:
Meas1
Measurement 1 or All (best three measurements)
FVC-Graph
:
Yes
yes; no
Flow Graph
No
yes; no
Diagnosis
:
Yes
1
yes; no
Lung Age
:
No
2
yes: no
PEF (l/min)
:
No
yes; no
MENU
CHOICE
Notes:
1
Diagnosis: The diagnosis statement (printed or not printed) is in accordance with the defined `normals` (see next
page and Section 2).
2
Lung Age: The lung age (printed or not printed) is the calculated nominal age of the patient`s lungs according to
the results obtained and the defined `normals` (see next page and Section 2). The value given is
calculated using the actual FEV1 value against predicted FEV1 value for the age of the patient. No value
will be given if the patient age is less than 18 years or if poor FEV1 test results have been obtained.