35-ml test volume
Refer to the illustration below:
1.
Move the syringe plunger to the 35-ml line.
2.
Drill a small hole (for example, 9/64”) through the syringe and the plunger shaft, at a
location between the plunger and the top of the syringe.
3.
Insert a rod or bolt (for example, a 6-32 screw) through the hole so that the plunger
cannot move, creating a constant volume in the syringe of 35 ml ± 2 ml.
4.
Secure the rod or bolt so that it cannot fall out of the hole.
--
-- 10
--
-- 20
--
-- 30
--
-- 40
--
-- 50
--
-- 60
ml