Service Manual
Disassembly and reassembly
53
REASSEMBLY:
(1) Clean all remnants of adhesive from the upper and lower cradle cases, and from the I/
O panel if you are reusing it.
(2) Apply a new adhesive gasket to the I/O panel.
Note: On USB-equipped cradles, keep the USB port cover out of the way during
reassembly.
(3) Connect the I/O panel to J1 on the Power board.
(4) Install the Power board into the slots in the upper housing.
(5) Set the I/O panel in place, but do not remove the protective film from the adhesive
gasket until the lower cradle housing is installed.
Removing the interconnect board
1.
Remove the four screws securing the Interconnect board to the upper housing and lift
the Interconnect board from the housing.
Removing the latch
1.
Remove the four screws holding the latch bracket, and remove the bracket.
Note
An early version of the Interconnect board included a photosensor attached by a
4-wire cable. If this photosensor and cable are present, clip the cables close to
the board and remove the cable and photosensor.
Reassembly: 4 inch lbs
Interconnect board
Reassembly: 4 inch lbs
Latch bracket screws (4)
Alignment slot
Alignment pinhole
Содержание Propaq LT 802LT0N
Страница 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Страница 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...