Service Manual
Disassembly and reassembly
27
Connectors
Zero insertion force (ZIF) flex cable connectors
ZIF connectors use a sliding outer piece that latches and unlatches to secure and release
the flex cable. ZIF cables cannot be successfully connected or disconnected without
properly unlatching and latching the sliding outer piece.
To disconnect a ZIF cable
1.
Using a suitable tool (paper clip, small flat-blade screwdriver, or small needle-nose
pliers), slide the latching piece of the connector away from the connector body and
toward the cable, releasing the cable from the connector.
2.
Remove the cable.
To connect a ZIF cable
1.
Slide the latching piece of the connector body away from the body.
Note
All connectors are keyed or labeled to facilitate proper connection. Take care to
correctly align all connector halves before attempting to connect them.